Device evaluated by manufacturer: the device was returned for analysis.The device was received in two sections as a result of a break in the hypotube.A visual and tactile examination identified a complete break in the hypotube of the device.The break was located at 255 mm distal to the strain relief.The hypotube was also kinked at more than one location.No other issues were noted with the hypotube that may have contributed to the complaint incident.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Due to the condition of the returned device it was not possible to inflate the balloon in the normal fashion, due to the break in the hypotube.As a result, an inflation aid was attached to the distal section of the hypotube break.This facilitated the connection of the encore inflation unit which allowed the balloon to be inflated.The returned device was attached to an encore inflation unit and positive pressure was applied, the balloon was inflated to its rate of burst pressure (rbp) without issue.A vacuum was then applied.The inflation device was verified at the rbp, before and after use with a pressure gauge.This inflation to rate of burst pressure was repeated three times with no leaks or drop in pressure noted.No issues were identified with balloon inflation or the balloon material.A visual and microscopic examination identified no damage to the tip, markerbands or blades of the device.All blades were present and fully bonded to the balloon material.No other issues were identified during the product analysis.
|