ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 29MECJ-502
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2019
Event Type  Injury  
Event Description
On (b)(6) 2019, a 29mm masters valve was selected for implant.During an internal rotation of the valve, while the device was attached with sutures, the valve broke.The device was exchanged for another size and implanted, doubling the procedure and extra corporal circulation time.
 
Manufacturer Narrative
The reported event of the valve "breaking" could not be confirmed.Both leaflets were properly inserted in the orifice and were able to fully open and close.No damage was noted to the valve.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8733043
MDR Text Key149143628
Report Number2648612-2019-00049
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734006545
UDI-Public05414734006545
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/05/2023
Device Model Number29MECJ-502
Device Catalogue Number29MECJ-502
Device Lot Number6623968
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2019
Was the Report Sent to FDA? No
Date Manufacturer Received08/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-