• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0313
Device Problems Signal Artifact/Noise (1036); Detachment of Device or Device Component (2907); Unintended Application Program Shut Down (4032)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, they had a short study.Technical support logged into the customer's computer and confirmed that instead of 48 hours, the study was only 23:50 hours and there were also a number of "ignores" in the study because the capsule fell off.The recorder worked correctly during the previous procedure.There was no patient and user harm and a repeat procedure was necessary.
 
Manufacturer Narrative
Evaluation summary: this report is based on information provided by medtronic investigation personnel.The product sample was not returned to the medtronic laboratory; however, a study graph was provided by the customer for analysis.The customer reported they had a short study.The reported condition was confirmed.The investigation found that the reported condition was due to bravo system communication failure.The investigation isolated the failure to the bravo communication system, but a cause was not identified.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
161 cheshire lane, suite 100
plymouth, MN 55441
7632104064
MDR Report Key8733304
MDR Text Key149159723
Report Number9710107-2019-00282
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101361688
UDI-Public07290101361688
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/03/2020
Device Model NumberFGS-0313
Device Catalogue NumberFGS-0313
Device Lot Number44431Q
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/04/2019
Initial Date FDA Received06/26/2019
Supplement Dates Manufacturer Received06/30/2019
Supplement Dates FDA Received07/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age45 YR
Patient Weight64
-
-