|
Model Number DOUBLECUT, 4.0MM X 13CM |
Device Problem
Peeled/Delaminated (1454)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 06/14/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
|
Event Description
|
It was reported during a shoulder arthroplasty procedure the ar-8400obe oval burr, and ar-8400dc double cut were leaving metal shavings in the shoulder post acromioplasty.The rep stated it is unknown if only the ar-8400obe was producing shavings, or if both devices were.The debris being produced were between the size of powder and shavings.The case was completed by using suction on the ar-8400dc to remove the shavings.The rep stated it seemed as though all shavings were removed.However, it cannot be confirmed if all shavings were fully suctioned out of the joint.No other burrs were opened.The speed of the console was set at 6200 rpm.Suction on the handpiece was set at 450 mmhg, and the disposable was in reverse.No outflow tubing was used, and suction was on through handpiece during acromioplasty.The bone quality was medium.
|
|
Manufacturer Narrative
|
The complaint could not be confirmed.Visual evaluation revealed no surface damage to the inner diameter of the outer hub, or any damage on the corresponding location at the distal end of the inner tuber that would indicate a site of metal debris production.Material analysis showed the returned ar-8400dc met all as-received specifications.No problem found.
|
|
Search Alerts/Recalls
|
|
|