Model Number 801188
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/04/2019
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) was unable to verify that the epgs would not calibrate but did hear that the stepper motor was loud and kept adjusting.He replaced the epgs.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) would not always calibrate and had a clicking sound.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.The service repair technician (srt) was unable to verify the reported complaint.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed the electronic patient gas system (epgs) fraction of inspired oxygen (fio2) and liters per minute (l/min) control knobs to make normal mechanical noises when performing calibration, flow testing and when controlled manually.No other noises were observed during evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8740229
MDR Text Key149380313
Report Number1828100-2019-00341
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)110317
Combination Product (y/n)N
PMA/PMN Number
K163531
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/04/2019
Initial Date FDA Received06/27/2019
Supplement Dates Manufacturer Received08/12/2019
09/10/2019
Supplement Dates FDA Received09/03/2019
09/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-