Model Number 801188 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) was unable to verify that the epgs would not calibrate but did hear that the stepper motor was loud and kept adjusting.He replaced the epgs.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) would not always calibrate and had a clicking sound.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) was unable to verify the reported complaint.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the electronic patient gas system (epgs) fraction of inspired oxygen (fio2) and liters per minute (l/min) control knobs to make normal mechanical noises when performing calibration, flow testing and when controlled manually.No other noises were observed during evaluation.
|
|