Model Number 3228
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Therapeutic Effects, Unexpected (2099); Inadequate Pain Relief (2388)
Event Date 06/06/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided when the evaluation is completed.
 
Event Description
It was reported the patient complained of not receiving effective stimulation therapy coverage from the scs system.A company representative met with the patient and attempted to reprogram the patient's scs system, but the patient reported only getting unintended rib stimulation.An x-ray was taken and confirmed the scs lead had migrated.Surgical intervention may be taken to address the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for evaluation.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Follow up information received identified the patient's scs lead was revised and moved to the t9 space.Postoperative, all painful areas were covered with adequate pain relief.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8740540
MDR Text Key149390032
Report Number1627487-2019-07491
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017253
UDI-Public05415067017253
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/07/2020
Device Model Number3228
Device Lot Number6716381
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/06/2019
Initial Date FDA Received06/27/2019
Supplement Dates Manufacturer Received08/09/2019
Supplement Dates FDA Received08/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-