ZIMMER BIOMET, INC. ECHO FX 7MM STD FEMORAL; PROSTHESIS, HIP Back to Search Results
Catalog Number 12-151307
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported during the distribution process packaging damage with sterility barrier potentially compromised was identified.No patient or surgical involvement.No further information available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported packaging damage event was confirmed via visual examination.It was identified the packaging was damaged and/or opened at an unknown time.Device history record (dhr) review was performed with no related manufacturing deviations or anomalies identified.Root cause is unknown.The likely condition of the product when leaving zimmer biomet was conforming to specifications.It cannot be determined when and where the sterile cavity was opened.Therefore, the root cause of the reported issue cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information available at the time of this reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO FX 7MM STD FEMORAL
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8743544
MDR Text Key150878845
Report Number0001825034-2019-02765
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K063614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number12-151307
Device Lot Number726230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2019
Was the Report Sent to FDA? No
Date Manufacturer Received10/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-