Catalog Number 904760
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdrs were filed in association with this event.Please see: 0001825034-2019-02792, 0001825034-2019-02793, 0001825034-2019-02794, 0001825034-2019-02799, 0001825034-2019-02800, 0001825034-2019-02801, 0001825034-2019-02802, 0001825034-2019-02803, 0001825034-2019-02804, 0001825034-2019-02806, 0001825034-2019-02807.Concomitant medical products: 908646, washerloc cortical screw 46mm, lot 305460.906513, tunneloc fixation, lot 465065.904781, standard ezloc, lot 576530.904755, ziploop, lot p08647.904056, ez pass, lot 545450.904055, ez pass, lot 359600.904053, ez pass, lot 048410.900321, maxfire, lot 909090.904760, toggle drill, lot 699980.912031, juggerknot, lot p09905.912029, juggerknot, lot p08797.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported during the distribution process packaging damage with sterility barrier potentially compromised was identified.No patient or surgical involvement.No further information available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Udi : (b)(4).Reported packaging damage event was confirmed via visual examination.It was identified the packaging was damaged at an unknown time.Device history record (dhr) review was performed with no related manufacturing deviations or anomalies identified.The likely condition of the product when leaving zimmer biomet was conforming to specifications.A definite root cause of the reported issue cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information available at the time of this reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM CANN TOGGLE BUTTON DRILL
Type of Device
BIT, DRILL
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8743558
MDR Text Key204919150
Report Number0001825034-2019-02805
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 10/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number904760
Device Lot Number699980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/20/2019
Initial Date FDA Received06/28/2019
Supplement Dates Manufacturer Received10/14/2019
Supplement Dates FDA Received10/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-