Model Number SBC-501-100
Device Problems Leak/Splash (1354); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/03/2019
Event Type  malfunction  
Manufacturer Narrative
Patient's exact age is unknown; however, it was reported that the patient was over the age of 18.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a seal biopsy valve was used in a colonoscopy/esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the cap came off and the physician was sprayed with fluid.It is noted that there is not enough hold to seal the cap on.The procedure was completed with another of the same device.There was no serious injury nor adverse patient effects reported as a result of this event.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available; a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEAL BIOPSY VALVE
Type of Device
SEAL BIOPSY VALVE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
NOLATO CONTOUR INC
660 vandeberg street
,
baldwin WI 54002
Manufacturer Contact
carole morley
300 boston scientific way
,
marlborough, MA 01752
5086834015
MDR Report Key8745760
MDR Text Key149638769
Report Number3005099803-2019-03282
Device Sequence Number1
Product Code OCX
UDI-Device Identifier08714729981619
UDI-Public08714729981619
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSBC-501-100
Device Catalogue Number2235-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/03/2019
Initial Date FDA Received06/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-