• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER; SWAN-GANZ CATHETER Back to Search Results
Model Number 774F75
Device Problem Material Rupture (1546)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2019
Event Type  malfunction  
Manufacturer Narrative
One catheter with detached monoject 1.5 cc limited volume syringe was returned for evaluation.The balloon was found to be ruptured.The ruptured edges were not able to match up.All through lumens were patent without any leakage or occlusion.No other visible damage or abnormality to the catheter body or returned syringe was observed.Visual examination was performed under microscope at 20x magnification.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon rupture was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is noted in the ifu to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz catheter ruptured during use.The customer commented that the volume injected into the balloon was 1.5 ml when the rupture occurred.There were no patient complications reported.Patient demographic information requested but unavailable.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key8746050
MDR Text Key149681315
Report Number2015691-2019-02373
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/27/2020
Device Model Number774F75
Device Catalogue Number774F75
Device Lot Number61708870
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2019
Initial Date Manufacturer Received 06/07/2019
Initial Date FDA Received06/28/2019
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received01/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-