Device evaluated by mfr.: the device was returned for analysis.Unit returned with its original pouch.Overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The device was slightly kinked at the distal tip.On the other hand, the device was inspected under high magnification at a distance of 8 inches from the distal tip and it was found that the device was a little scratched.The overall length, outer diameter (od) of distal tip, od of middle of the device and od of proximal section were within specifications.
|
Reportable based on device analysis completed on 12jun2019.It was reported that guidewire deformation occurred and guide wire insertion difficulties were encountered.During preparation of a 035/260/1 (bx/1) amplatz super stiff guide wire, it was noted that the taper was frayed 8 inches from the distal end and it was unable to load in the sheath.The procedure was completed with a new wire.No patient complications were reported.However, returned device analysis revealed guide wire coating peeled/sheared.
|