BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 101025
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: the event occurred on an unknown date in (b)(6) 2019.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that approximately two hours into treatment with a gambro cartridge set, an external blood leakage was observed at the venous injection port.It was reported a disconnection of the line to the venous port was observed.Treatment was discontinued with blood restitution.It was reported the blood loss was less than 150 ml.The cassette was replaced to continue treatment.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photo identified an external blood leak; however, it was not possible to detect the exact location where the leak occurred.The reported condition was verified.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTRIDGE
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE - TIJUANA BAJA
tijuana baja california
MDR Report Key8749839
MDR Text Key182990186
Report Number8030638-2019-00009
Device Sequence Number1
Product Code FJK
UDI-Device Identifier07332414007836
UDI-Public(01)07332414007836
Combination Product (y/n)N
PMA/PMN Number
K070414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2021
Device Catalogue Number101025
Device Lot Number1000210317
Was Device Available for Evaluation? No
Date Manufacturer Received07/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
-
-