BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Catalog Number 101025 |
Device Problems
Disconnection (1171); Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: the event occurred on an unknown date in (b)(6) 2019.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that approximately two hours into treatment with a gambro cartridge set, an external blood leakage was observed at the venous injection port.It was reported a disconnection of the line to the venous port was observed.Treatment was discontinued with blood restitution.It was reported the blood loss was less than 150 ml.The cassette was replaced to continue treatment.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photo identified an external blood leak; however, it was not possible to detect the exact location where the leak occurred.The reported condition was verified.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|