MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 031-33J
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 06/11/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported that the connecting part assembly to flowmeter was unstable and not properly connected when screwed.Therefore, a new unit was used instead.No patient involvement reported.
 
Event Description
It was reported that the connecting part assembly to flowmeter was unstable and not properly connected when screwed.Therefore, a new unit was used instead.No patient involvement reported.
 
Manufacturer Narrative
(b)(4).One (1) unit of 031-33j "nebulizer adaptor 033, sterile, japanese" was received for analysis.Signs of use are observed since the sample was not received on its original package.Also, during visual inspection it was observed that the adaptor doesn't spin freely.No other issues were found.Oxygen entrainment testing was performed and during the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing but the testing failed due to the unstable condition.After the testing finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection wear was found on the internal tabs.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
MDR Report Key8750449
MDR Text Key149771692
Report Number3004365956-2019-00182
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date11/03/2021
Device Catalogue Number031-33J
Device Lot Number74L1600507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2019
Date Manufacturer Received07/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-