RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
Device Problem
Display Difficult to Read (1181)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2017.Date of report: 02july2019.The manufacturer's field service engineer (fse) was dispatched to the site and confirmed the reported problem.The manufacturer's field service engineer replaced the lcd display to resolve the reported issue.The cracked top enclosure was also replaced.
|
|
Event Description
|
The customer reported a ventilator with a display that was too dark, not adjustable.The customer reported that there was no patient involvement.
|
|
Manufacturer Narrative
|
Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|