• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON VENTILATOR TUBING SET,LONG LENGTH; CIRCUIT, BREATHING (W CONNECTO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON VENTILATOR TUBING SET,LONG LENGTH; CIRCUIT, BREATHING (W CONNECTO Back to Search Results
Catalog Number 1613
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 06/13/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Received one (1) 1613, ventilator tubing set, long length, for investigation.Upon receipt the tubing was visually inspected for any signs of abuse/misuse/ damage.During visual inspection, a slit was visually identified on the column side of circuit, approximately 10 inches below the wye.The slit is approximately 3/4 of an inch in length.The device history record of batch number 74m1800028 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.The dhr shows that the product was assembled and inspected according to specifications.Based on the investigation performed, the reported complaint was confirmed.A significant "slit" was discovered in the blue tubing causing profuse leaking.The slit is very linear , which would suggest the tubing came into dangerous contact with a sharp object.Although the complaint has been confirmed, a root cause could not be established.
 
Event Description
The complaint is reported as: "crack found in circuit prior to set-up during leak test." no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON VENTILATOR TUBING SET,LONG LENGTH
Type of Device
CIRCUIT, BREATHING (W CONNECTO
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key8755854
MDR Text Key149928855
Report Number3004365956-2019-00180
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number1613
Device Lot Number74M1800028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2019
Initial Date Manufacturer Received 06/13/2019
Initial Date FDA Received07/02/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-