RESPIRONICS CALIFORNIA, INC BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
Device Problem
Defective Device (2588)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2017.Date of report: 03july2019.The reported issue of unit alarm failure was confirmed by a philips technician.The manufacturer's field service engineer (fse) replaced main and display case assy and the blower and flow valve assy.Ran performance verification and all testing passed.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
|
|
Event Description
|
The customer reports that the unit alarms and that the ventilator is inoperable.The unit was being used on a patient at the time the reported issue occurred; however, there was no patient harm.
|
|