• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD Back to Search Results
Model Number 4542
Device Problems High impedance (1291); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/22/2019
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this lead was surgically abandoned due to lead fracture.The lead exhibited high pacing impedance and when checking, the fluoroscope revealed the fracture.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
GUIDANT PUERTO RICO, B.V.
no. 12, road 698
,
dorado PR 00646 3311
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8758904
MDR Text Key149971062
Report Number2124215-2019-13045
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/21/2007
Device Model Number4542
Device Catalogue Number4542
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/22/2019
Initial Date FDA Received07/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/21/2005
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age82 YR
-
-