Model Number 9770600Y
Device Problems Failure to Charge (1085); Battery Problem (2885)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, unit will not hold a charge.It works when plugged in but shuts down when removed from the wall.
 
Event Description
Per biomed, unit will not hold a charge.It works when plugged in but shuts down when removed from the wall.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the battery will not hold a charge is confirmed.The scanner shuts off prematurely after ac power is removed.The root cause of the reported issue is an internal li-ion battery failure.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the battery will not hold a charge is confirmed.The scanner shuts off prematurely after ac power is removed.The root cause of the reported issue is an internal li-ion battery failure.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, unit will not hold a charge.It works when plugged in but shuts down when removed from the wall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8761412
MDR Text Key150258613
Report Number3006260740-2019-01887
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138287
UDI-Public(01)00801741138287
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770600Y
Device Catalogue Number9770600Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2019
Event Location Hospital
Initial Date Manufacturer Received 06/13/2019
Initial Date FDA Received07/03/2019
Supplement Dates Manufacturer Received10/02/2019
10/11/2019
Supplement Dates FDA Received10/17/2019
11/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number2243072-10/11/2019-013-C