MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 UNKNOWN HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Unspecified Infection (1930); Inflammation (1932); Injury (2348); Osteolysis (2377); No Code Available (3191)
Event Date 06/18/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the summit stem with pinnacle metal on metal implanted on 2005 presented with significant metallosis and osteolysis.It was also indicated that there was loosening of the stem at bone to implant interface.The proximal stem had 1 cm bone loss medial and from greater trochanter.It was explanted easily.The acetabulum had a significant defect superior.Pinnacle 300 series was still well fixed and difficult to remove.Interoperative cultures showed moderate signs of infection, so the surgeon decided to removed all components and implant a prostalac spacer.Doi: (b)(6) 2005; dor: (b)(6) 2019, right hip.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR CUP
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8771156
MDR Text Key150392102
Report Number1818910-2019-97804
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Date Manufacturer Received07/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
-
-