Manufacturer's investigation conclusion: the centrimag motor used during the reported event was not returned for analysis.Multiple attempts for product return were made but the motor was not returned.As a result, the reported event could not be confirmed and the root cause could not be conclusively determined.Reports of similar events will continue to be tracked and monitored.The 2nd generation centrimag system operating manual section 3.1.5-"centrimag and pedivas blood pumps" states "the blood flow is dependent upon the amount of blood entering the pump, the pump speed (rpm), the extracorporeal circuit resistance, and drainage and return blood pressures.The relationship between pressure and flow rate as a function of rpm can be seen in figure 16 for a complete centrimag vas circuit (including the pump, tubing, and one representative set of cannulae that may be used with the pump)." this document also has an emergency/troubleshooting section for the 2nd generation console.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console, and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.No further information was provided.The manufacturer is closing the file on this event.
|