After intensive investigation under the microscope, the (b)(6) medical center found the avmid packs / kits products have been the source of fibers that were discovered in the eyes of several pts on day 1 and 1 week.Providers notes: fibers on the field within this package are seen on the sterile blue drape (source) loose and freely attaches to anything that lands on the field.Paper towels (possible source) - when rubbed or touch by gloves, lets out fibers onto the gloves 4x4, qtips cta - (least likely source, as these are difficult to pull off and rub off) - not free, weaved tightly.Blue basins are massively infected by particles only seen under microscope and with perfect lighting using a bubble of bss fluid to see it floating i the droplet.Otherwise not able to see with naked eye or in the surgical eye at time of surgery with brown or blue iris background.Black dust particles, red fibers/lint, green fibers/lint, lots of white dust / debris.Syringes (open, are not individually packed) found in the basins have fibers inside of them, inside the syringe barrel.This was compared to syringes that were individually packaged bd brand name not from the avmed pack, which were crystal clear inside.Fda safety report id# (b)(4).
|