The device was returned to avante health solutions (ahs) for evaluation.Visual inspection identified that the bottom case had not been changed during previous repair; however, there is no known possibility of this causing overheating of the device nor patient harm.The correct ahs cable was connected to the fetal transducer at the time of the previous repair.Testing was performed by keeping the device plugged in for 48 hours straight which determined that there was no heat generating from the device.The cable worked as expected; however, there was a torn membrane.A definitive root cause for the reported patient burn cannot be determined; however, it is possible that the patient had a negative reaction to type of gel being used during the use of the device.No further action is required.Avante health solutions will continue to monitor this type of event.
|