MAUDE Adverse Event Report: INTEGRA LIFESCIENCES(IRELAND) CUSA EXCEL 36KHZ STRAIGHT HANDPIECE; ULTRASONIC SURGICAL PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES(IRELAND) CUSA EXCEL 36KHZ STRAIGHT HANDPIECE; ULTRASONIC SURGICAL PRODUCTS Back to Search Results
Catalog Number C2602
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A sales representative reported on behalf of the customer that the c2602 cusa excel 36khz straight handpiece failed during surgery on an unspecified date.The handpiece alarm was going off after pressing test.The sales representative disconnected the handpiece and put it all back together again to make sure the technician didn't make an error and got the same problem.They pulled another handpiece to check to see if that would work and it did with no alarms.There was no patient injury reported.Additional information has been requested.
 
Manufacturer Narrative
The product was returned for evaluation and the technician found that the handpiece amplitude dropped down to 10%.The device history record (dhr) review showed no abnormalities that could be associated with the complaint incident.The reported failure was confirmed.The cause of the complaint incident was due to a faulty transducer.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSA EXCEL 36KHZ STRAIGHT HANDPIECE
Type of Device
ULTRASONIC SURGICAL PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
MDR Report Key8775694
MDR Text Key150603095
Report Number3006697299-2019-00077
Device Sequence Number1
Product Code LBK
Combination Product (y/n)N
PMA/PMN Number
K141674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC2602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2019
Initial Date Manufacturer Received 06/14/2019
Initial Date FDA Received07/10/2019
Supplement Dates Manufacturer Received07/12/2019
Supplement Dates FDA Received07/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-