Model Number FT4 G3
Device Problem Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/17/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The initial reporter stated that they received discrepant results for one patient sample tested with the elecsys tsh assay, elecsys ft3 iii, and the elecsys ft4 iii assay on a cobas 8000 e 801 module.The sample also had a discrepant value when tested with the elecsys ft4 ii assay on a second e 801 analyzer that was used for investigation.The results measured on the reporter's analyzer were reported outside of the laboratory.This medwatch will apply to the ft4iii assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the tsh assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the ft3 assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the ft4ii assay.The sample was tested at the customer site on (b)(6) 2019.The sample was repeated on an abbott architect analyzer.The sample was also provided for investigation, where it was tested on a second e 801 analyzer on (b)(6) 2019.No adverse events were alleged to have occurred with the patient.The serial number of the reporter's e 801 analyzer is (b)(4).The serial number of the e 801 analyzer used for investigation is (b)(4).Ft4iii reagent lot number 380330, with an expiration date of december 2019 was used on this analyzer.
 
Manufacturer Narrative
The customer sent the patient sample to the manufacturer for investigation.The customer's ft3, ft4 ii, and tsh results were reproduced.The differences of the tsh and ft4 values, generated with the different types of analyzers, are caused by differences of the setups of the assays, the antibodies used and differences of the standardization materials and procedures used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key8779939
MDR Text Key150757741
Report Number1823260-2019-02510
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 08/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age54 YR
-
-