Catalog Number M003SZAS45210 |
Device Problem
Premature Separation (4045)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/18/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that after transferring the stent (subject device) into the microcatheter and while advancing the stent, only the delivery wire protruded out of the microcatheter tip and the stent had prematurely deployed during use and remained stuck inside the microcatheter.The physician replaced it with new devices and continued the procedure without clinical consequences to the patient.
|
|
Manufacturer Narrative
|
The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.During visual inspection, the stent was returned inside the microcatheter with an additional stent.The stent delivery wire was kinked/bent.The stent was deformed.No other anomalies were noted.Functional testing could not be performed due to the condition of the returned device.Based on device analysis, it is likely that the stent delivery wire slipped through the stent due to the presence of the second stent in the microcatheter.The damage noted to the stent delivery wire and stent is also likely to have been due to the presence of the second stent.Therefore, an assignable cause of procedural factors will be assigned to this investigation.
|
|
Event Description
|
It was reported that after transferring the stent (subject device) into the microcatheter and while advancing the stent, only the delivery wire protruded out of the microcatheter tip and the stent had prematurely deployed during use and remained stuck inside the microcatheter.The physician replaced it with new devices and continued the procedure without clinical consequences to the patient.
|
|
Search Alerts/Recalls
|