ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71940-01
Device Problem Device Displays Incorrect Message (2591)
Patient Problems Hypoglycemia (1912); Nausea (1970); Paralysis (1997); Shaking/Tremors (2515)
Event Date 06/15/2019
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer received the error message, " scan again in 10 minutes," when scanning the adc freestyle libre sensor.On (b)(6) 2019, the customer experienced trembles, nausea, and "corps (muscle) paralysis." the customer was treated by a non-hcp with oral glucose.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, the case will be re-opened, and a physical investigation will be performed.
 
Event Description
A customer received the error message, " scan again in 10 minutes," when scanning the adc freestyle libre sensor.On (b)(6)2019, the customer experienced trembles, nausea, and "corps (muscle) paralysis." the customer was treated by a non-hcp with oral glucose.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
tara williamson
1360 south loop road
alameda, CA 94502-7001
5108644472
MDR Report Key8787302
MDR Text Key150947631
Report Number2954323-2019-05401
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71940-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/17/2019
Initial Date FDA Received07/12/2019
Supplement Dates Manufacturer Received01/23/2020
Supplement Dates FDA Received02/19/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age33 YR
Patient Weight120
-
-