• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND BTK TURBOHAWK; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND BTK TURBOHAWK; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number THS-SX-C
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 07/12/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the physician used a turbohawk device with a non-medtronic 6fr sheath and a 0.014 300cm nitrex wire to treat a 150mm plaque, 90% stenotic lesion in the patient¿s mid superficial femoral artery (sfa).The vessel diameter was reported as 6mm with severe tortuosity.The ifu was followed during preparation, procedure and post procedure.The vessel was not pre-dilated.It was reported that the atherectomy was performed.The physician encountered high resistance while attempting to remove the device.After 2/3 attempts, the device was removed.After inspection, it was noticed that the nosecone was detached.Attempts to retrieve it with a snare were unsuccessful so the physician decided to cage the nosecone with a self-expanding stent.The vessel was post dilated.The physician was satisfied with the angiographic results.No patient injury was reported.
 
Manufacturer Narrative
Additional information: the nose cone detached in the mid sfa.The snare used was a non-medtronic device.A medtronic everflex stent was used to cage the detached nose cone to the vessel wall.There were no patient complications reported.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis: the hawkone device was returned for analysis.No ancillary device or images were received with the device.The cutter driver was not returned with the hawkone.Inspection of the hawkone revealed the distal assembly contained biological debris throughout.The distal rotating tip was detached from the assembly just distal to the flush window; the rotating tip was not returned.The tecothane remained intact.Multiple bends were noted throughout the housing, the housing was bent in a ¿c¿ shape.The inner coils at the proximal end of the device were bent and separated.The inner coils at the distal end were stretched and pulled and were protruding out from the distal end.The cutter remained attached and was approximately 2.0cm distal to the cutter window.Microscopic inspection of the guidewire lumen revealed a zipper tear throughout the assembly.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BTK TURBOHAWK
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key8788596
MDR Text Key150956573
Report Number9612164-2019-02846
Device Sequence Number1
Product Code MCW
UDI-Device Identifier00643169968226
UDI-Public00643169968226
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/24/2022
Device Catalogue NumberTHS-SX-C
Device Lot Number0009657185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/12/2019
Initial Date FDA Received07/15/2019
Supplement Dates Manufacturer Received07/15/2019
09/14/2019
Supplement Dates FDA Received07/26/2019
09/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-