• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PROVISIONAL ARTICULAR SURFACE SIZE YELLOW 14 MM HEIGHT; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PROVISIONAL ARTICULAR SURFACE SIZE YELLOW 14 MM HEIGHT; TEMPLATE Back to Search Results
Catalog Number 00597103014
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 01/14/2019
Event Type  malfunction  
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.(b)(4).Multiple mdr's were filed for this event.Please see associated: 0001822565 - 2019 - 00394, 0001822565 - 2019 - 02953, 0001822565 - 2019 - 02959, 0001822565 - 2019 - 02960, 0001822565 - 2019 - 02963, 0001822565 - 2019 - 02964, 0001822565 - 2019 - 02966, 0001822565 - 2019 - 02970, 0001822565 - 2019 - 02971, 0001822565 - 2019 - 02972, 0001822565 - 2019 - 02974, 0001822565 - 2019 - 02976.Concomitant medical products: lot #: 62109025, description: provisional, articular surface, item #: 00597103012; 60267319, provisional, articular surface, 00597103014; 60702359, provisional, articular surface, 00597103014; 62069849, provisional, articular surface, 00597103014; 60037678, provisional, articular surface, 00597103017; 60718197, provisional, articular surface, 00597103017; 61723504, provisional, articular surface, 00597103017; 60711711, provisional, articular surface, 00597104014; 61716917, provisional, articular surface, 00597104014; 60695901, provisional, articular surface, 00597104017; 60711697, provisional, articular surface, 00597105012; 61704885, provisional, articular surface, 00597105014; 61661748, provisional, articular surface, 00597105017; 60695888, provisional, articular surface, 00597103010; 61354292, provisional, articular surface, 00597103010; 62245587, provisional, articular surface, 00597103010; 63631356, provisional, articular surface, 00597103010; 63631356, provisional, articular surface, 00597103010; 63944239, provisional, articular surface, 00597103010; 63963834, provisional, articular surface, 00597103010; 60263008, provisional, articular surface, 00597103012; 60692099, provisional, articular surface, 00597103012; 63613874, provisional, articular surface, 00597103012; 63661447, provisional, articular surface, 00597103012; 63944252, provisional, articular surface, 00597103012; 63485270, provisional, articular surface, 00597103014; 63620103, provisional, articular surface, 00597103014; 63649634, provisional, articular surface, 00597103014; 63492795, provisional, articular surface, 00597103017; 63624551, provisional, articular surface, 00597103017; 63637863, provisional, articular surface, 00597103017; 63950874, provisional, articular surface, 00597103017; 60711685, provisional, articular surface, 00597104010; 63544460, provisional, articular surface, 00597104010; 63564705, provisional, articular surface, 00597104010; 63649636, provisional, articular surface, 00597104010; 63707471, provisional, articular surface, 00597104010; 63907831, provisional, articular surface, 00597104010; 63923219, provisional, articular surface, 00597104010; 60723046, provisional, articular surface, 00597104012; 63707472, provisional, articular surface, 00597104012; 63729498, provisional, articular surface, 00597104012; 63909551, provisional, articular surface, 00597104012; 63950876, provisional, articular surface, 00597104012; 63963833, provisional, articular surface, 00597104012; 64098766, provisional, articular surface, 00597104012; 62106300, provisional, articular surface, 00597104014; 63624555, provisional, articular surface, 00597104014; 63655336, provisional, articular surface, 00597104014; 63667496, provisional, articular surface, 00597104014; 60263016, provisional, articular surface, 00597104017; 62117392, provisional, articular surface, 00597104017; 63505873, provisional, articular surface, 00597104017; 63612752, provisional, articular surface, 00597104017; 63624556, provisional, articular surface, 00597104017; 63644014, provisional, articular surface, 00597104017; 60706801, provisional, articular surface, 00597105010; 63492799, provisional, articular surface, 00597105010; 63624557, provisional, articular surface, 00597105010; 63644015, provisional, articular surface, 00597105010; 63661449, provisional, articular surface, 00597105010; 63707475, provisional, articular surface, 00597105010; 64135297, provisional, articular surface, 00597105010; 60503279, provisional, articular surface, 00597105012; 63485275, provisional, articular surface, 00597105012; 63649638, provisional, articular surface, 00597105012; 63673401, provisional, articular surface, 00597105012; 63944254, provisional, articular surface, 00597105012; 64135298, provisional, articular surface, 00597105012; 60692115, provisional, articular surface, 00597105014; 61579539, provisional, articular surface, 00597105014; 63492801, provisional, articular surface, 00597105014; 63583780, provisional, articular surface, 00597105014; 63661450, provisional, articular surface, 00597105014; 63661450, provisional, articular surface, 00597105014; 62750675, provisional, articular surface, 00597105017; 63169365, provisional, articular surface, 00597105017; 63490729, provisional, articular surface, 00597105017; 63579633, provisional, articular surface, 00597105017; 63581233, provisional, articular surface, 00597105017.The event occurred in (b)(6).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
The instruments were returned fractured and/or worn as identified in inventory.No patient or surgical involvement reported.No further information is available at this time.
 
Event Description
No further information is available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The visual evaluation of returned provisional articular surfaces shows signs of repeated use and all of them have been cracked/fractured.The dimensions for these instruments were conforming to prints where measured.This information confirms the reported event.Review of the device history record with receiving inspection report identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue was due to repeated use of these instruments for more than 5-years of field life; which causes wear and tear to the instrument and led to the fracture/cracks.Per package insert (87-6203-999-22 instrument/ provisional use, care and sterilization) inspect all instruments/provisional carefully prior to each use.If damage or wear is noted that may compromise the function of the instrument, do not use the device and contact your zimmer representative for a replacement.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROVISIONAL ARTICULAR SURFACE SIZE YELLOW 14 MM HEIGHT
Type of Device
TEMPLATE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8790588
MDR Text Key151109776
Report Number0001822565-2019-02952
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00597103014
Device Lot Number60267319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/08/2019
Initial Date FDA Received07/15/2019
Supplement Dates Manufacturer Received10/22/2019
Supplement Dates FDA Received10/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-