A new airseal ifs system was to be used for surgical procedure and the standard process was followed to prepare the airseal ifs system for the transanal minimally invasive surgery (tamis) procedure.Prior to using, error code: "valve error call service." company service was called from or room.Followed service representative's directions.This was a major fault and not operational.Surgeon had to use co2 directly from the laparoscopic machine to insufflate the rectum.This worked, but was less than optimal.The airseal ifs system was brought to the biomedical engineering department.The manufacturer is sending a new replacement system via overnight delivery.Manufacturer response for airseal ifs system surgiquest, airseal ifs system surgiquest (per site reporter).Replacement of the new non-functioning ifs system.
|