The reported event was confirmed.A temporary pacing electrode catheter was returned.A single clamp was attached to a black tail.The clamp was removed and it was found that the black tail was broken.The clamp was attempted to be placed on the non-broken tail with no success.The connection was attempted again and the clamp was able to be connected.A potential root cause could be due to "user error/mistake".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "1.Inspect the sterile package carefully for damage during transit or storage.Do not use the catheter if the package 2.Visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage.3.In case of catheters with a balloon, under sterile conditions, remove the protective sheath and inflate the balloon with 1.5 ml of air or carbon dioxide.Use the inflation syringe included in the package.Completely deflate the balloon after the test.".
|