• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC Back to Search Results
Model Number P3700E000009
Device Problem Unintended System Motion (1430)
Patient Problem No Patient Involvement (2645)
Event Date 07/04/2019
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the right siderail board needed to be replaced.Per the hillrom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance (pm).Pm will minimize downtime due to excessive wear.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the right siderail board to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from the hillrom service technician stating the bed hi/lo was going up on its own.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
TABLE, OBSTETRIC
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key8798202
MDR Text Key151322463
Report Number1824206-2019-00275
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP3700E000009
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/04/2019
Initial Date FDA Received07/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-