• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPEEDBAND SUPERVIEW SUPER 7; LIGATOR, HEMORRHOIDAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPEEDBAND SUPERVIEW SUPER 7; LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number M00542251
Device Problems Failure to Fire (2610); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the stomach during an endoscopic variceal band ligation (evl) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the suture cut off and the bands failed to deploy.The procedure was completed with another seedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Problem code 2610 for the reportable issue of bands failed to deploy.Problem code 2907 for the reportable issue of suture detached.Investigation results received one speedband superview super 7 with the ligator head for analysis.It was noticed that the crimp was present on the trip wire and the trip wire was secured in the handle assembly slot when received.A visual examination of the ligator head found five bands present which were moved out of their original positions and one band was caught under the other bands.It was also noticed that the ligator teeth were bent.The suture was broken; was attached to the ligator head, and the other portion was attached to the trip wire loop.A functional evaluation was performed by rotating the handle knob 180 degrees, an audible click was heard and indents were felt.No visible issue was noted with the handle assembly.Based on the evaluation of the returned complaint device, these failures are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.It is most likely that the ligator head teeth were damaged due to handling and manipulation of the device during the procedure.Once the ligator head teeth are damaged, the suture can be detached from its position on the ligator head and this condition could have impacted the bands deployment activity, moving bands without being properly deployed and contributing to the reported issues.Also, the suture can tangle with the bands which can result in difficulty to rotate the handle and continued attempts to deploy the bands can result in suture breakage.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.Correction: e1 (initial reporter zip/post code).
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the stomach during an endoscopic variceal band ligation (evl) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the suture cut off and the bands failed to deploy.The procedure was completed with another seedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the stomach during an endoscopic variceal band ligation (evl) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the suture cut off and the bands failed to deploy.The procedure was completed with another seedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Block h6: problem code 2610 for the reportable issue of bands failed to deploy.Problem code 2907 for the reportable issue of suture detached.Block h10: investigation results received one speedband superview super 7 with the ligator head for analysis.It was noticed that the crimp was present on the trip wire and the trip wire was secured in the handle assembly slot when received.A visual examination of the ligator head found five bands present which were moved out of their original positions and one band was caught under the other bands.It was also noticed that the ligator teeth were bent.The suture was broken; was attached to the ligator head, and the other portion was attached to the trip wire loop.A functional evaluation was performed by rotating the handle knob 180 degrees, an audible click was heard and indents were felt.No visible issue was noted with the handle assembly.Based on the evaluation of the returned complaint device, these failures are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.It is most likely that the ligator head teeth were damaged due to handling and manipulation of the device during the procedure.Once the ligator head teeth are damaged, the suture can be detached from its position on the ligator head and this condition could have impacted the bands deployment activity, moving bands without being properly deployed and contributing to the reported issues.Also, the suture can tangle with the bands which can result in difficulty to rotate the handle and continued attempts to deploy the bands can result in suture breakage.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8798214
MDR Text Key151299796
Report Number3005099803-2019-03618
Device Sequence Number1
Product Code FHN
UDI-Device Identifier08714729201960
UDI-Public08714729201960
Combination Product (y/n)N
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/23/2019
Device Model NumberM00542251
Device Catalogue Number4225
Device Lot Number0022831035
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2019
Date Manufacturer Received08/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age75 YR
Patient Weight70
-
-