According to the reporter, during set-up for hysteroscopy, on the end of the scope, there was a metal piece that was like a burr that was sharp.And the lens looks like it has a line through it.Replaced with another scope and it worked fine.There was no patient involvement.
|
Additional information: evaluation summary: one device was received for evaluation.The returned sample did not meet specification as received by medtronic.A visual inspection was performed and showed the outer tube is broken near to the welding seam between the outer tube and the main body, the distal end is damaged, there is moisture inside the optical system and a crack in an optical component inside the optical system.The reported condition was confirmed.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned.¿ these words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|