The device history record confirms that the device met all material, assembly and performance specifications.Visual inspection was performed on the returned device and it was observed that the stent was contained within the sheath and the stent was deformed.The distal tip of the sheath was damaged.During functional evaluation, the sheath was flushed, and the stent delivery wire and stent were unable to be advanced through the sheath.The stent delivery wire and stent were retracted back through the sheath gently and the stent was noted to be broken/ fractured with the distal end remaining in the sheath.It is probable that the tip of the sheath was damaged during insertion into the hub of the microcatheter causing the reported and analyzed issues; therefore an assignable cause of procedural factors will be assigned to this investigation.
|