• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number A-TCSE-DF
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem ST Segment Elevation (2059)
Event Date 06/28/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was not reviewed as the batch number was not available.Based on the information received, the cause of the reported st segment elevation was procedure related.
 
Event Description
Related manufacturing reference: 2030404-2019-00056.During a pulmonary vein isolation ablation procedure st segment elevation occurred.While ablating the left sided veins with the tacticath se catheter an alarm went off on the cool point irrigation pump.The heparin saline bag was empty and needed to be replaced to continue the procedure.The tubing set was not disconnected from the catheter in the patient and the stop cock was inadvertently left in the on position to the patient and air entered the left atria during priming.Global segment elevation occurred along with hypotension and asystole.Approximately two minutes later the patient returned to normal rhythm and the st elevation returned to baseline.The procedure was not continued.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8802122
MDR Text Key151420787
Report Number3005334138-2019-00386
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA-TCSE-DF
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
COOL POINT¿ IRRIGATION PUMP TUBING SET
Patient Outcome(s) Other;
-
-