• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN SAPPHIRE ANALYZER; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN SAPPHIRE ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 08H00-01
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
All available patient information has been included.No additional patient information is available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer reported a disagreement in wbc results between instruments when processing on the cell-dyn sapphire.The sapphire result was 0.295 and repeat on the ruby was 5.490.Another additional sapphire result was 0.24 and retest was 8.0 on the ruby.After troubleshooting the sapphire results matched the ruby results.No impact to patient management was reported.
 
Manufacturer Narrative
An investigation was performed for the customer issue and included a review of the complaint text, troubleshooting, a review of product historical data, and a review of product labeling.The issue resolved by replacing the wbc mixpot and peristaltic pump #3 for wbc staging.Review of product historical data did not identify any adverse trends or abnormal complaint activity.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation a product deficiency was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN SAPPHIRE ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key8810174
MDR Text Key151720027
Report Number2919069-2019-00086
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740016616
UDI-Public00380740016616
Combination Product (y/n)N
PMA/PMN Number
K051215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H00-01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/22/2019
Initial Date FDA Received07/21/2019
Supplement Dates Manufacturer Received08/29/2019
Supplement Dates FDA Received08/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-