This is a supplemental report.The subject maj-1500 was returned to olympus medical systems corp.(omsc) for evaluation.One of two lock levers at the equipment side connector of maj-1500 was broken.Based on the similar case, the lock lever likely broke due to degradation and repeated loading during long time use.The instruction manual provides of maj-15000 provides warnings and how to inspect the device prior to use as follows.There were no further details provided.If significant additional information is received, this report will be supplemented.Warning over time, the connecting tube will deteriorate because it is subject to wear with each use.If any of the following are found with the connecting tube, do not use it and replace it with a new one.Connecting a defective tube could prevent the effectiveness of the cleaning and high-level disinfection process.Inspection move the lock levers of the connecting tube to make sure that they function properly and are not broken.
|