It was reported that the effect of the ablation could not be observed.The patient underwent an ablation procedure for atrial fibrillation.Upon unpacking the intellanav oi ablation catheter it was observed to be kinked and when bending in both directions an audible crack was heard and the catheter did not bend far enough.During the procedure in the left atrium it was felt that though there was irrigation, the catheter temperature was unusually low at 21 degrees c, and the impedance seemed to change little; the effect of the ablation could not be felt.The device was in power control mode.Signal noise was also present, which was improved by improving the power supply around the equipment.Magnetic positioning allowed for small cardiac vein isolation and the procedure to be successfully completed without patient complications.
|