Model Number DTBA1D4 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Abscess (1690); Unspecified Infection (1930); Weakness (2145)
|
Event Date 07/02/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: 5076-52 lead, implanted: (b)(6)2010.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient's system was removed due to systemic infection.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The device was returned and screening analysis was performed, but no issue was identified requiring full analysis.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The patient is a participant in the post approval clinical surveillance product surveillance registry.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that physician thought that the source of the infection was likely an abscess in the muscles in the pelvis region.The infection was considered to be secondary and unrelated to the procedure or products used.
|
|
Manufacturer Narrative
|
Corrected: a2, a4, a5a, a5b, b2, b5, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient's cardiac resynchronization therapy defibrillator (crt-d) system was removed due to the patient having septicemia and bacteremia.Vegetation was observed on the leads and blood cultures tested positive for enterococcus faecalis.A left iliopsoas muscle abscess was also observed.It was noted that the patient experienced generalized weakness.A replacement system was placed a few days later with an antibacterial absorbable envelope.The patient is a participant in the post approval clinical surveillance product surveillance registry.No further patient complications have been reported as a result of this event.
|
|