• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND ARCHITECT TOTAL PSA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND ARCHITECT TOTAL PSA Back to Search Results
Catalog Number 07K70-27
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
This report is being filed on an international product list 7k70, that has a similar us product distributed in the us, list 6c06.An evaluation is in process.A follow up report will be submitted when the evaluation is completed.
 
Event Description
The account generated a false elevated architect total psa of 140.9 ng/ml on a patient that did not have a diagnosis of prostate cancer.The sample repeated 28.858 (neat) but 144.29 ng/ml (1:5 diluted).No impact to patient management was reported.Race and ethnicity of the patient is unknown.
 
Manufacturer Narrative
Lot updated from unknown to 95445fn00.A review of result logs for instrument serial (b)(4).Was performed and confirms the customers observation for elevated total psa results for one sample draw.A new sample draw was tested within close proximity on the same instrument using the same reagent lot and a result of 28.858 ng/ml was obtained.Controls were running as expected.It is possible that the initial elevated results were due to a specimen issue.Additionally, the reviewed results show no atypical signal profiles for the 2 samples processed.The ticket searches determined normal complaint activity for the likely cause lot.Complaint trending report review determined that there is no adverse trend for the product for the complaint issue.Instrument data analytics was reviewed which determined that the patient median result for the complaint lot is comparable with other lots in the field and confirms no systemic issue for the lot.No customer returns were available for evaluation.A review of the product quality history for the lot number using search of the corrective and preventive actions system did not identify issues associated with the customer observation.Labeling was reviewed which adequately addresses the current issue.No product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TOTAL PSA
Type of Device
TOTAL PSA
Manufacturer (Section D)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI  NA
MDR Report Key8828883
MDR Text Key160373582
Report Number3008344661-2019-00091
Device Sequence Number1
Product Code MTF
Combination Product (y/n)N
PMA/PMN Number
P910007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/09/2020
Device Catalogue Number07K70-27
Device Lot Number95445FN00
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/02/2019
Initial Date FDA Received07/25/2019
Supplement Dates Manufacturer Received09/06/2019
Supplement Dates FDA Received09/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARCHITECT I2000SR LIST 03M74-01; ARCHITECT I2000SR LIST 03M74-01; SERIAL (B)(4); SERIAL (B)(4)
Patient Age63 YR
-
-