The following information was reported to gore: on (b)(6) 2019 a patient was undergoing treatment of a popliteal aneurysm with a gore® viabahn® endoprosthesis.Access was obtained via the contralateral femoral artery, in an up and over approach, with a short 7fr sheath over a.014" 300cm spartacus wire.The device was advanced to the target location.When deployment was attempted the endoprosthesis curled up like a snake and the deployment line was wrapped around the device, terminating expansion.The device had opened approximately 1-2cm.The device was removed from the vessel without apparent damage, however the procedure was converted to open bypass.The patient did well following the bypass procedure.
|
H.6.Results code 2: 213: the engineering evaluation stated the following: the endoprosthesis, part of the delivery catheter and part of the deployment line were returned.The hub of the delivery catheter was not returned.A guidewire was returned, this was not evaluated as it is not a gore device.There was approximately 138 cm of deployment line returned with two single fibers coming from the end that measured approximately 14 and 8 cm long.The dual lumen was stretched in two locations.The first was approximately 12 cm from the hub end of the delivery catheter for about 4 cm.The second was approximately 34 cm from the hub end of the delivery catheter and was stretched for 23 cm.The distal shaft, upon which the endoprosthesis is mounted, was broken at the transition.Part of the distal shaft was visible within the transition.The endoprosthesis was partially expanded approximately 5 cm with approximately 9 cm of the endoprosthesis still constrained by the constraining line.There was body delamination as well as outwardly flared struts in the body of the endoprosthesis.Deployment was able to be continued with traction on the deployment line at the endoprosthesis.Based on the device examination performed, no manufacturing anomalies were identified which could be definitively attributed to the event.
|