Model Number DL-344-10-K
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/27/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).The explanted lead has not yet been returned to neuropace for analysis.However, review of the neurostimulator ecog and lead impedances for the left depth lead were suggestive of a potential lead break.
 
Event Description
A review of the patient's ecog data indicated lost ecog signal and demonstrated high impedances, suggestive of a possible lead break on the left depth lead.The patient reported hitting her head on her bed, during the winter of 2018 however no other contributing factors were reported by the patient or treating center.The patient underwent a procedure on (b)(6) 2019 to remove the lead and implant a new left depth lead no complications were reported during the procedure and the signals and impedances for the newly implanted lead were within normal limits.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key8829472
MDR Text Key152473565
Report Number3004426659-2019-00030
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005014
UDI-Public010085554700501417191027
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDL-344-10-K
Device Catalogue Number1007607
Device Lot Number22889-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age24 YR
-
-