Model Number FT4 G2
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation found for the event that from the information provided, a general reagent issue most likely can be excluded.To the differences of the ft4 values generated from different types of analyzers: it needs to be taken into account that assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods and the standardization methodology used.There was no follow up/corrective action.No devices were returned.This device is not labeled for single use and is not reprocessed or reused.
 
Event Description
This report summarizes 1 malfunction events.Questionable low results were generated by elecsys ft4 ii assay.The event involved a total of 1 patient on a cobas 6000 e 602 module.The provided patient's age was (b)(6) years.The patient was a female.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 II ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
mannheim (baden-wurttemberg), 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
indianapolis, IN 46250
MDR Report Key8830182
MDR Text Key152669418
Report Number1823260-2019-90210
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberFT4 G2
Device Lot NumberASKU
Type of Device Usage N
Patient Sequence Number1
-
-