Model Number INFANT FLOW SIPAP
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/10/2019
Event Type  Injury  
Manufacturer Narrative
Results of investigation: engineers found that the internal power cable was loose, causing the unit to lose power and could not start.The cable was reconnected and reinforcement of the equipment were normal.
 
Event Description
It was reported to vyaire medical that the customer was not able to "boot" during assisted breathing.There was no report of any patient harm associated with this reported event, however, the patient was irritated at the time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Type of Device
VENTILATOR, NON-CONTINUOUS
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key8832806
MDR Text Key152380980
Report Number2021710-2019-10576
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberINFANT FLOW SIPAP
Device Catalogue Number27408-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-