Catalog Number 5100060001
Device Problem Failure to Auto Stop (2938)
Patient Problem Abrasion (1689)
Event Date 07/02/2019
Event Type  Injury  
Manufacturer Narrative
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
 
Event Description
It was reported that during a cranial procedure, the perforator bit did not stop and tore the dura.It was also reported that there was a few minutes delay to repair the dura.It was further reported that the procedure was completed successfully.
 
Manufacturer Narrative
The perforator product reported involved with this event was returned for evaluation and the disengagement mechanism within the perforator was tested and functioned as intended.The reported failure of failure to disengage could not be confirmed.The ifu #5100-060-700 rev.F provides instructions on how to test the disengagement function of the perforator prior to use.The ifu also warns: use extreme caution when drilling in conditions such as: ¿ bone that may vary in consistency and/or thickness greater than 1 mm; ¿ adherent dura; ¿ high intracranial pressure; ¿ other abnormalities in the area of penetration.The perforator may cut or nick the dura or brain.
 
Event Description
It was reported that during a cranial procedure, the perforator bit did not stop and tore the dura.It was also reported that there was a few minutes delay to repair the dura.It was further reported that the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYPHR DISPOSABLE CRANIAL PERFORATOR, LARGE 14/11MM
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-IRELAND
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
Manufacturer Contact
una barry
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
214532900
MDR Report Key8836650
MDR Text Key152499512
Report Number0001811755-2019-02576
Device Sequence Number1
Product Code HBF
UDI-Device Identifier04546540716224
UDI-Public04546540716224
Combination Product (y/n)N
Reporter Country CodeSE
PMA/PMN Number
K082010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2022
Device Catalogue Number5100060001
Device Lot Number19078027
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-