Model Number 781341
Device Problem Temperature Problem (3022)
Patient Problems Burn, Thermal (2530); Partial thickness (Second Degree) Burn (2694)
Event Date 07/19/2019
Event Type  Injury  
Manufacturer Narrative
Based on the provided information and tests performed on site, there is no indication of a malfunction of the mri system or coil used that contributed to the injury.The blistering on both thumb and buttocks is explained by the direct contact between the patient¿s thumb and buttock.No other contributing factors were identified.
 
Event Description
Philips received a report from a customer related to a heating incident with the posterior coil on an ingenia 1.5t system.A patient received a 3 centimeter blister on both the buttock and thumb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIA 1.5T
Type of Device
SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
dusty leppert
veenpluis 4-6
p.o. box 10.000
best 5680 -DA
NL   5680 DA
MDR Report Key8839381
MDR Text Key152602954
Report Number3003768277-2019-00060
Device Sequence Number1
Product Code LNH
UDI-Device Identifier00884838055322
UDI-Public(01)00884838055322
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K110151
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 07/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number781341
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
Patient Weight86
-
-