• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 12.2MM ONE STEP REAMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 12.2MM ONE STEP REAMER Back to Search Results
Catalog Number 14-442002
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported during the distribution process packaging damage with sterility barrier potentially compromised was identified.No patient or surgical involvement.No further information available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Packaging was not returned for investigation.Only the device was returned.The reported issue of sterile packaging damage cannot be confirmed.Review of the device history records identified no deviations or anomalies during manufacturing.This device is used for treatment.The likely condition of the product when leaving zimmer biomet was conforming to specifications.A definite root cause of the reported issue cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information available at the time of this reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12.2MM ONE STEP REAMER
Type of Device
REAMER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8840806
MDR Text Key152636883
Report Number0001825034-2019-03289
Device Sequence Number1
Product Code HTO
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2014
Device Catalogue Number14-442002
Device Lot Number101047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2019
Was the Report Sent to FDA? No
Date Manufacturer Received10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-