• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179702000S
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/08/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Hospital - (b)(6).Date of complaint - 8/7/19.Made aware - 8/7/19.Present in theatre - yes.Delay in procedure - approx 5-10min.Impact to patient - none.Product code (1)- 179712600, lot: gm4196503.Product code (2) - 179702000s, qty = 4, lot = unknown.Am i in possession - awaiting sterilisation.What happened - point of torque wrench shaft was spinning in set screw head, we had to change to alternative instruments.Evidence of wear at hex of shaft.4 set screws stripped out leaving sharp metal debris.This complaint involves five (5) devices.
 
Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination of the returned device found the threads on the set screw were completely torn off.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.With the information provided, a definitive root cause for the torn threads on the single-inner set screw cannot be determined.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE-INNER SETSCREW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8841183
MDR Text Key175725402
Report Number1526439-2019-51892
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034265063
UDI-Public(01)10705034265063
Combination Product (y/n)N
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179702000S
Device Catalogue Number179702000S
Device Lot Number243369
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2019
Date Manufacturer Received09/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-