The manufacturer was informed that during an unspecified procedure, the biomed at the user facility reported that when using the unit, a patient experienced a shock at the grounding pad.There was no information if there was any additional harm done to the patient due to being shocked.Additionally, the biomed indicated that the patient plate was expired and that the unit was set to 10watt of power in cut mode.The staff only brought the referenced unit for testing and the biomed noted that he could not determine the cause of the reported "patient was shocked at the patient plate" as there was no patient plate to test with the unit.The unit was tested and the outputs of the unit were within the required range.
|