• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK F/G WINGSPAN STENT SYSTEM 3.0 X 20MM; STENT, INTRACRANIAL NEUROVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK F/G WINGSPAN STENT SYSTEM 3.0 X 20MM; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Catalog Number M003WE0300200
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/27/2019
Event Type  malfunction  
Manufacturer Narrative
The device history record review confirms that the device met all material, assembly, and performanc e specifications.Visual and microscopic examination of the returned device found that the distal and proximal stent delivery catheter was noted to be kinked.The proximal stabilizer was kinked.The rhv was loosened and it was noted that the stabilizer was broken.The stent was found to be partially deployed.Functional testing could not be performed due to the anomalies found on the device.Based on the analysis, the reported event was confirmed.Information available indicated that the device was confirmed to be in good condition prior to use.Based on the investigation results and available information, it is possible that the anatomical factors present during the clinical procedure caused the damage to the device and the reported defects.Therefore, a cause of procedural factors and error usage have been assigned to this investigation.
 
Event Description
Based on the investigation of the returned product, the stent stabilizer was found broken inside patient.The stent was found to be partially deployed.There were no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
F/G WINGSPAN STENT SYSTEM 3.0 X 20MM
Type of Device
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key8843965
MDR Text Key152764506
Report Number3008881809-2019-00220
Device Sequence Number1
Product Code NJE
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
H050001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/19/2021
Device Catalogue NumberM003WE0300200
Device Lot Number20679190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/26/2019
Initial Date FDA Received07/30/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient Weight70
-
-